Fluke Corporation MET/CAL Procedure ============================================================================= INSTRUMENT: Fluke 8062A: (1 year) CAL VER /5100,5205,8842 DATE: 11-Feb-95 AUTHOR: Fluke Corporation REVISION: $Revision: 2.4 $ ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 45 NUMBER OF LINES: 211 CONFIGURATION: Fluke 5100B CONFIGURATION: Fluke 5205A CONFIGURATION: Fluke 8842A ============================================================================= # # Source: # Fluke 8062A instruction manual (P/N 632695, March 1982, Rev. 2, 3/93) # # Compatibility: # MET/CAL 4.0 or later # # Subprocedures: # None # # Required Files: # 51_COMP4.bmp # 51_V8062.bmp # 51HV8062.bmp # 51_A8062.bmp # # System Specifications: # TUR calculation is based on specification interval of the accuracy file. # The default 5100B accuracy file contains 6 month specs. # # Fluke makes no warranty, expressed or implied, as to the fitness # or suitability of this procedure in the customer's application. # # The 6 month specifications of the 5100B are used in TUR computations. # STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- N P F V 1.002 ASK+ K 1.003 HEAD EQUIPMENT SETUP 1.004 DISP [32] WARNING 1.004 DISP HIGH VOLTAGE is used or exposed during the performance 1.004 DISP of this calibration. DEATH ON CONTACT may result if 1.004 DISP personnel fail to observe safety precautions. 1.005 DISP As UUT is turned ON, verify that all display 1.005 DISP segments momentarily light up. 1.006 EVAL Did the UUT display test pass? 2.001 DISP Ambient temperature: 18C - 28C. 2.001 DISP Relative humidity: less than 80%. 2.001 DISP Allow the UUT 30 min. to stabilize. 2.002 HEAD {DC VOLTAGE TESTS} 2.003 PIC 51_V8062 2.004 HEAD DC VOLTAGE TESTS: {200mV Range} 2.005 5100 200 190.00mV 0.11U 2W 3.001 5100 200 -190.00mV 0.11U 2W 4.001 HEAD DC VOLTAGE TESTS: {2V Range} 4.002 5100 2 1.9000V 0.0011U 2W 5.001 HEAD DC VOLTAGE TESTS: {20V Range} 5.002 5100 20 19.000V 0.015U 2W 6.001 HEAD DC VOLTAGE TESTS: {200V Range} 6.002 5100 200 190.00V 0.15U 2W 7.001 HEAD DC VOLTAGE TESTS: {1000V Range} 7.002 5100 1000 1000.0V 0.9U 2W 8.001 HEAD {AC VOLTAGE TESTS} 8.002 HEAD AC VOLTAGE TESTS: {200mV Range} 8.003 5100 200 100.00mV 0.60U 50H 2W 9.001 5100 200 100.00mV 1.40U 10kH 2W 10.001 HEAD AC VOLTAGE TESTS: {2V Range} 10.002 5100 2 1.0000V 0.0060U 200H 2W 11.001 5100 2 1.0000V 0.0070U 1kH 2W 12.001 5100 2 1.0000V 0.0070U 10kH 2W 13.001 5100 2 1.0000V 0.0140U 30kH 2W 14.001 5100 2 0.1000V 0.0015U 200H 2W 15.001 5100 2 0.1000V 0.0050U 30kH 2W 16.001 HEAD AC VOLTAGE TESTS: {20V Range} 16.002 5100 20 10.000V 0.060U 200H 2W 17.001 5100 20 10.000V 0.520U 10kH 2W 18.001 5100 20 10.000V 0.540U 30kH 2W 19.001 HEAD AC VOLTAGE TESTS: {200V Range} 19.002 5100 200 100.00V 0.60U 200H 2W 20.001 5100 200 100.00V 5.20U 10kH 2W 21.001 DISP Disconnect the 5100B from the UUT. 21.002 PIC 51HV8062 21.003 5205 200 100.00V 5.40U 30kH 2W 22.001 HEAD AC VOLTAGE TESTS: {750V Range} 22.002 5205 750 750.0V 16.0U 100H 2W 23.001 5205 750 750.0V 16.0U 1kH 2W 24.001 DISP Disconnect the 5205A from the UUT. 24.002 ASK- N U 24.003 HEAD RESISTANCE TESTS (Test Lead Compensation) 24.004 PIC 51_COMP4 24.005 IEEE [@8842]G5[I] 24.006 MATH MEM1 = MEM - 1100 24.007 JMPF 24.010 24.008 DISP Select FRONT INPUTS on the 8842A. 24.009 JMP 24.005 24.010 RESF 100Z S 2W 24.011 8842 100Z N 4W 24.012 MATH M[1] = MEM 24.013 RESF 1kZ S 2W 24.014 8842 1kZ N 4W 24.015 MATH M[2] = MEM 24.016 RESF 10kZ S 2W 24.017 8842 10kZ N 4W 24.018 MATH M[3] = MEM 24.019 RESF * S 24.020 HEAD {RESISTANCE TESTS} 24.021 HEAD RESISTANCE TEST: {200 Ohm Range} 24.022 DISP Short the V Ohm S and COMMON inputs together. 24.023 RNG 200Z 24.024 ASK- U 24.025 ACC 0.00Z TOL 24.026 MEMI Enter UUT reading in ohms: 24.027 MEME 24.028 MEMC 200 Z +0.04U 25.001 DISP Remove the short. 25.002 PIC 51_V8062 25.003 RESF 100.00Z S 2W 25.004 ASK+ N U 25.005 MATH MEM = M[1] # 8842A 90 day accuracy, 200 Ohm Range 25.006 ACC 100 Z 0.007% 0.004U 25.007 MEMI Enter UUT reading in Ohms: 25.008 MEME 25.009 MEMC 200 Z 0.14U 26.001 ASK- U 26.002 HEAD RESISTANCE TESTS: {2 kOhm Range} 26.003 RNG 2kZ 26.004 RESF 1.0000kZ S 2W 26.005 ASK+ U 26.006 MATH MEM = M[2] # 8842A 90 day accuracy, 2 kOhm Range 26.007 ACC 1 kZ .005% .00003U 26.008 MEMI Enter UUT reading in kilohms: 26.009 MEME 26.010 MEMC 2 kZ 0.0012U 27.001 ASK- U 27.002 HEAD RESISTANCE TESTS: {20 kOhm Range} 27.003 RNG 20kZ 27.004 RESF 10.000kZ S 2W 27.005 ASK+ U 27.006 MATH MEM = M[3] # 8842A 90 day accuracy, 20 kOhm Range 27.007 ACC 10 kZ .005% .0003U 27.008 MEMI Enter UUT reading in kilohms: 27.009 MEME 27.010 MEMC 20 kZ 0.012U 28.001 HEAD RESISTANCE TESTS: {200 kOhm Range} 28.002 RESF 200 100.00kZ 0.12U 2W 29.001 HEAD RESISTANCE TESTS: {MOhm Range} 29.002 DISP Set UUT to the MOhm range 29.003 RESF 1.0000MZ 0.0022U 2W 30.001 RESF 10.00MZ 0.05U 2W 31.001 HEAD {CONTINUITY TEST} 31.002 RNG 2kZ 31.003 DISP Press the -><-))) button (just below the display) once. 31.004 EVAL Is the -><- symbol visable on the display? 32.001 ASK- U 32.002 RESF 0.1000kZ S 2W 32.003 ASK+ U # 8842A 90 day accuracy, 200 Ohm Range 32.004 TOL 0.003% 1.4e-6U 32.005 ACC 0.1000kZ TOL 32.006 MATH MEM = M[1] / 1000 32.007 MEMI Enter UUT reading in kilohms: 32.008 MEME 32.009 MEMC kZ 0.0003U 33.001 DISP Press the -><- button again to enable the audible tone. 33.002 ASK- U 33.003 RESF 100Z S 2W 33.004 ASK+ U 33.005 EVAL Is the audible tone sounding? 34.001 EVAL Does the UUT Indicate "OL" ? Is the audible tone OFF ? 35.001 HEAD {DIODE TESTS} 35.002 DISP Select the Diode function 35.003 EVAL Does the UUT Indicate "OL" ? 36.001 ASK- U 36.002 RESF 1.0000kZ S 2W 36.003 ASK+ U # 8842A 90 day accuracy, 2 kOhm Range 36.004 ACC 1.0000kZ 0.003% 8e-6U 36.005 MATH MEM1 = M[2] 36.006 MEMI Enter UUT reading in kilohms: 36.007 MEME 36.008 MEMC kZ 0.1000U 37.001 HEAD {DC CURRENT TESTS} 37.002 PIC 51_A8062 37.003 HEAD DC CURRENT TESTS: {200uA Range} 37.004 5100 200 190.00uA 0.59U 2W 38.001 5100 200 -190.00uA 0.59U 2W 39.001 HEAD DC CURRENT TESTS: {2mA Range} 39.002 5100 2 1.9000mA 0.0059U 2W 40.001 HEAD DC CURRENT TESTS: {20mA Range} 40.002 5100 20 19.000mA 0.059U 2W 41.001 HEAD DC CURRENT TESTS: {200mA Range} 41.002 5100 200 190.00mA 1.35U 2W 42.001 HEAD DC CURRENT TESTS: {2000mA Range} 42.002 5100 2000 1900.0mA 13.5U 2W 43.001 5100 2000 -1900.0mA 13.5U 2W 44.001 HEAD {AC CURRENT TESTS} 44.002 HEAD AC CURRENT TESTS: {20mA Range} 44.003 5100 20 19.000mA 0.152U 1kH 2W 45.001 END